All posts by Simon

HP 8753C Network Analyzer: an Avantek fix

Still continuing on the repairs of the HP 8753C. No luck with the Watkins-Johnson YTO, maybe one could get it to work after some severe modification of the phase lock assembly, but this is not my intention. So I rather try another YTO, a Avantek part, purchased for 25 USD.

Some changes are necessary – at least the dampening of the YIG needs some work. Set the 8753 to the source tune mode, and determined the capacitor value to get stable current regulation (in the source tune service mode, the PLL is open).
From about 100-120 nF onwards, in parallel to a 100 nF-2.2 kOhm network, perfectly stable. Added 200 nF, ceramic capacitors, directly at the YTO coil, and modified the R-C network on the A11 assembly a bit (replaced the 1 k resistor with a 2.2 k).

Here a quick schematic, the voltage regulators, and a few Zener diodes to protect the YTO driver. The main modifications are:

*1 cut a trace on the A11 assemly, re-route to the current sense resistor, to allow for some adjustment (sure you can use fixed resistor, once the necessary current has been found.

*2 replace the current sense resistor – most YTOs have 50 mA/GHz current requirement, about double of the original part.

*replace the YTO driver transistor, with a part that can be more suitably mounted on a heatsink, or provide some heatsink to the existing TO-3, if you can manage.

*adjust the R-C/C network across the YTO main coil, good values are 100-200 nF in series with 1-5 kOhm, in parallel with 10-250 nF. Check for clean and fast current regulation, with the PLL open (source tune mode in the service menu).

After the preliminary checks (adjust the current of the YTO for about 4 MHz output, with the source tune mode set to 300 kHz), immediate success – there are absolutely no PLL issues, phase lock is established smoothly.

There is a 3 dB pad, the YTO has plenty of power, and it will help to reduce any frequency pull.

Below, to adjust the pretune DAC, you need to remove the EEPROM write protection, this picture from the web is quite useful.

At first try – the pre-tune test passed!!

Even better – all internal test passed!

Some check at 1 GHz, the spectrum looks clean, there are no parasitic oscillations or sidebands.

The analog bus of the 8753 series, input 16, is quite useful to check the tuning voltage and YTO status. It needs to be a line (the voltage is directly proportional to the YTO coil current).

To test the unit, measured a 1.9 GHz bandpass. All good.

Also the GPIB interface and plot function, working well.

After all the electronics, a little bit of work on the lathe – the cable that connects the test set to the VNA had broken screws, two new long screws fabricated – I used a bit thinner round stock, to make the task easier. Note that the cable is still available from Keysight, EUR 238 a piece.

Remaining task – to add the options 010 and 002, time domain analysis, and harmonic analysis, both are software options and can be added with a secret key code….

HP 8753C Network Analyzer: spare YTOs

Not giving up on the HP 8753C Network Analyzer repairs. Especially, because these units are really top class in terms of specifications and serviceability, easy to use, and well handled by GPIB bus software. Not sure why you would by and later model, if you have a working 8753C or similar model.

Some study of the YIG oscillators used in the 8753 series analyzers, while these are nothing special neither in output or tuning range (3.8-6.8 GHz, and a quick test on the source assembly shows that everything above 8 dBm seems to drive the source mixer to saturation), they have a ~23 mA/GHz tuning current, much less than the 50 mA/GHz (20 MHz/mA) of many industry standard and common YTOs. No idea why that is, a bit less power consumption, a bit less copper?

Looking around, found this marvelous Watkins-Johnson YTO, for just about 15 EURs. It was a bit dirty with wires badly soldered to it, but easily cleaned up.

Power is good, measured through a 6 dB attenuator, and pretty stable all over the range (won’t even need it up to 8 GHz).

a href=”https://www.simonsdialogs.com/wp-content/uploads/2020/02/8753-yig-replacement-power.jpg”>

There were no data on the power supply, so I needed to test it out. Heater voltage, seems to work well with 15 Volts, supply voltages -5 and +12 V give stable operation with sufficient margin (there seem to be internal voltage regulators).

The tuning current 50 mA/GHz, about double of the 23 mAh/Ghz of the HP YTO. So, there will be more heat dissipated, let’s do some calculations around the YTO driver. Located on the A11 board.

Clearly, the transistor will have to dissipate more heat. Do avoid any big changes, let’s adjust the current sense resistor to about half the value – resulting in about the same voltage drop per GHz.

The driver transistor, nothing special, a NPN TO-3 part, used without a heatsink.

There is no good space to fit a heatsink for TO-3 transistor, so I replaced the transistor. Used a BD245C, with amplification hfe of 40, well good enough. The SOT-93 case, it can easily be mounted on a piece of sheet metal (aluminum) to provide enough cooling.

Everything well insulated and mounted. Using the TO-3 screws to hold the transistor/heatsink assembly.

The current sense resistor, the key part for any YTO driver, needs to be low drift, low thermal coefficient. Otherwise, there will be all kinds of drift. HP use in some of their equipment sense resistors specified to 2 ppm/K, the best I could get is about 15 ppm/K with Dale RH-10. One day I will check their actual thermal coefficient. Now I just burned it in a bit, and selected one that looked perfectly stable with any load changes up to 1 Ampere.

The A11 board can be modified fairly easily, even reversibly – replace the sense transistor, from 40 Ohms to 20 Ohms, cut a trace (the current sense voltage to the opamps, close to the board connector – the via is handy to attach a wire). And a 18 Ohm/390 Ohm/200 Ohm -10 turn pot arrangement to set the proper currents.

Another small modification, the 1 k/100 network across the main coil – replace the 1 k resistor by 2.2 k.

The 8753C provides +15 V and -15 V at the YTO connector. Assembled a small power regulator board, to provide the necessary +12 V and -5 V.

Testing…

Several adjustments of the 10T trimmer, but there are issues – no stable lock below 3.3 MHz, and the Service Function 58 won’t do the pre-tune corrections. Note that you can use the source tune mode to monitor the YTO with the PLL disengaged. Ideal for adjustments of the YTO slope. Still no stable lock. Working reasonably stable at times, but all a bit temperamental.

Various lock issues, and the self tests won’t work well.

Some study of the manual, and quite some time spent to add more components, changing PLL filters, and so on. But no luck. Below 3.3 MHz, there is a also a gain change of the PLL, by Q10 FET, also modified a bit there, some improvement, but not as stable as it needs to be.

Maybe, some specific issue of this YTO, at least after detail study of the tuning currents, some magnetic hysteresis, or similar. Not an uncommon problem. Not all YTOs are suitable for fast, precise sweeping and phase-locked operation.

After all, let’s give it another try, with another YTO. Found this beauty for USD 25, a great AVANTEK YTO, ASF-8347M, with solid output power. Hermetically sealed, in a magnetically shielded casing.

HP 8753C Network Analyzer: all the dead YIG Oscillators

Having had the non-working HP 8753C on the bench for while, almost getting a bit desperate. The machine is in pristine condition, including the test set, and the CPU board issues were an easy fix. But workings source assemblies, or YTOs, are to no avail. If at all, they are way too expensive. Finally, even got a 2nd 8753C (this unit with option 006 and 002, and with a good display, good to have, in any case) – but same issues there – no signal, the YTO dead.

Here are the two YTOs, both not working at all.

Doing some search on the internet, seems that these oscillators have general lifetime issues, if you want to call it an issue after 30 years… I expect these to work 50 years and longer.

The 8753C series instruments used a variety of YIGs, all pretty similar, but different in their parts number. From hearsay, HP optimized the cost a bit and used these so call economy YIG oscillators. They are a bit suspicious, because they are not hermetically sealed.

The inner circuit, nothing to fix here, all put together with molybdenum substrate, gold wire, and a special transistor that doesn’t seem to provide enough gain any more. Sure I tried to rearrange the sphere (red arrow), and checked all the wires, but no apparent issue. It really seems to be a HF transistor issue inside the microcircuit, at the green arrow.

Be mindful when bidding for any of these 8753 units – they may have all dead or soon-to-die YTOs. And even during their prime time, when spares were still available from HP, the source assemblies 08753-60003 were only available as a whole unit, and certainly HP used to charge several 1000 USD per piece.

HP 8662A Synthesized Signal Generator: An almost new generator, with a pretty old power supply, and a serial number mystery

This is another one of the marvelous HP 8662A generators, I got it eventually for free because it had parts missing but otherwise, a good unit.

The missing parts were the mains filter, and the transformer for the OCXO. The filter, I just replaced it with a fairly standard commercial unit, should be good enough for now. The transformer it is a bit more tricky, because of the limited space and rate type. By luck and coincidence I found a spare in Japan. 25 Euros, fair enough.

Be aware that there are many versions of the input assemblies – and my unit has a 1982 vintage power supply, with a 1996 back plane. This is not going to work well, so I needed to study all the wires and connections. No wonder someone had given up on the repair job before.

The later version:

Mounted these parts – yet, I didn’t turn the unit on, because of a blown fuse. After some inspection of the power supply, the common issues. A few bad caps on the driver assembly, replaced with a pair of new 22 uF axial caps.

Other issues – one of the main switching transistors is dead – ordered some spare transistors, new old stock, from Greece, and took them to Japan during the last business trip to Germany… these are MJ16012 power transistors. 800 Volt, 15 Amp 175 Watts. I don’t trust any copies of these, but rather bought good old Motorola parts. 1986 vintage.

Good practice with the 8662A power supply repairs – checked the control assembly.

The oscillator is working, 40 kHz present, but no drive output – how come? Some quick checks clearly show issues around the U1 inverter, a defective logic gate. Someone must have tried to fix it before, but seems he didn’t have the proper tools and solder – all sticky and bad, so I cleaned it with alcohol.

Desoldered the U1 4049 CMOS, with quite some difficulty because the drill used was quite small, and difficult to desolder.

Some strange observation – all the versions and date codes. The power supply control assembly, maybe 1997 vintage. The power supply base assembly and low voltage circuit, 1982 vintage. The A7A3 power converter assembly, 1993 with some parts dating back to 1986…

The serial – 34xx would suggest 1960+34=1994, but this can’t be, because the frame was only made past 12/1995.

Checked two random assemblies – also these were made in 1997.

Probably fair to say, a 1997 unit, and someone tried to fix it some time after that with a 1982 supply. Why he changed the power supply baseboard, and the removed the oven transformer, no clue.

The defective gate of the control assembly – order a few new CMOS circuits. Plenty of them back home in Germany, but here, it is faster and cheaper to order them new from Thailand… in addition, ordered a few 4013 and so on just in case.

Lastly, this unit also has an attenuator missing. Not uncommon for the 8662A. It is a 33321-60028 65 dB 5 V attenuator. 5/40/20 dB elements. Maybe I will find one on eBay, but actually, I have not much use for it because this 8662A will be running at +10 dBm all the time as a reference LO for my VCO test rig and phase noise analyzer.

A precision current source: a mirror, and a TL431

There are many uses for a good current source, in particular, to drive a noise generator, Noise Source TWS-N15. Not much to write home about, but because of frequent requests, I am publishing the circuit here. It will work for small current from 2 or 3 mA up to 10 or 20 mA with no problem, and very little drift over temperature and time. For R, uses a good resistor. Input voltage can be up to 35 V, or even higher.

The big crash: Server failure

This blog is hosted by a professional provider, but the manuals archive (which needs quite a bit of storage), and other webpages, and my fileserver, is running on two machines, a Dell OptiPlex FX160 as the main, eco-efficient system (in Germany), and a Dell PowerEdge SC1425 with a Raid 1, 3 TB hard drive system as the backup, and currently my main system in Japan (where I am living on a temporary business assignment). Recently, the SC1425 failed, it just would not start up anymore. Power supply seems OK – likely, a severe issue. Checked all the memory and everything, but to no avail.

After fiddling around for about 2 hours, and still no success, I decided to order a new server – a new old server, Dell PowerEdge 850. Just about 35 Dollars used. Rather than 2x XEON processors, it has a Pentium D, 3.2 GHz Dual-Core. Plenty of power for a web- and fileserver.

A couple of days later, the unit arrived – removed the SATA Raid controller (running on Ubuntu with software Raid), and some BIOS settings (activate SATA, disable Keyboard error, enable boot from USB, default power up status is ON) plus BIOS Update. Also, reconfigured the router to make sure this machine will get all the HTTP requests.

A few tests – the harddrive is working fine, about 100 MB/s (sure there is a cache). The Raid 1 is up with no repairs or anything.

A quick check – also the web server is reachable.

I wouldn’t recommend a single PowerEdge for your super critical applications, but they are pretty good for the current cost, as long as you don’t mind the fan noise.

u-blox GPSDO: a simple, low cost, yet – high performance approach

There are many circuits around in the web, related to GPSDOs, and a more sophisticated design with a self-steering u-blox receiver has been published earlier here. Now I felt tempted to try an easier approach, without the hassle of precision references, operational amplifiers, DAC, and other devices that are great but high cost when you need to avoid noise and other complications.
Essentially, this design is a clean-up PLL, with some monitoring of the receiver, and the PLL health. All monitored by a simple 8 bit microcontroller, an ATMega8-16PU in this case.

We have some elements here, (1) the OCXO and amplifier, distribution amplifier – to provide the outputs, 4 in this case, and a good TTL level 10 MHz signal for the PLL, (2) a u-blox receiver, configured to provide either 5 Hz flashing in non-locked condition (no GPS reception, or no good reception), and 125 kHz, 50% duty cycle as a phase reference in locked condition, (3) the MCU, ATMega8, that is configuring the GPS received, providing a 125 kHz signal derived from the 10 MHz OCXO (the OCXO is used as the microprocessor clock – don’t introduce a new clock in such circuits, which will only lead to spurious signals!), (4) a 74HC86 that is used as a phase detector, and to convert the GPS output (a 3.3 V signal) to 5 V level.

That’s the OCXO and distribution amplifier…

The phase detector…

The controller and PLL filter – a simple two pole filter. It replaces all the expensive references, DACs and opamps of the more sophisticated designs. There is another small, faster filter to convert the phase angle to a voltage – converted by the 10 bit ADC of the ATMega8, 1 bit is about 4 ns.

The circuit full view…

Some first tests turned out well. Monitoring the OCXO phase with a scope…

To do a more thorough tests, without all my various test gear that it back in good old Germany, I used the 10 MHz to run another GPS receiver (after upconverting to a 26 MHz clock), then the NAV-CLOCK message can be used to report phase and drift. The short term stability of the OCXO is better than the GPS, as can be seen, but there is no long term drift – because the OCXO is now steered by the 1st GPS receiver via the PLL (XOR phase detector and loop filter).

The phase detection is done at 125 kHz, a convenient frequency for precise measurement, and high enough for filtering.

About 20 ns of jitter are clearly visible in the u-blox output, because it is running on a 48 MHz internal clock.

The circuit is running well, because of the few parts the cost is low and should be easy to reproduce. Let me know in case you need the ATMega code (written in GCC).

The display shows the phase angle, essentially, the duty cycle of the phase comparator output, the stability of the OCXO voltage (by a low pass algorithm), and the lock condition of the GPS (detected by measuring the frequency with timer0 of the ATMega8, and the INT0 interrupt at rising flank to reset the timer).

Phase noise is very small, there are no visible spurs (the lines seen on the screen relate to recalibration events of the analyzer rather than spurious signals, except those at +-125 kHz – at -90 dBc – probably you can get rid of these by better shielding and compartmentalization).

Sure there could be more sophisticated phase noise measurement, by analyzing the control voltage with a low frequency analyzer. I may proceed with such analysis these days but don’t expect to find much, anyway, would be best to fit the circuit to a shielded box first.

All in all, I believe this is a very workable solution that will give you great performance at lowest cost, and with little effort. Sure it will work with various types of OCXOs, the Trimble unit used is generally very good in terms of drift and phase noise. Be aware that some newer Trimble units aren’t all that good. The OCXO draws about 2 Amps at 12 Volts upon startup, but it is OK to start it with a current limited supply, at about 1 Amp, if you don’t want to overdesign the power supply.

GPSDO: a new 10 MHz distribution (and isolation!) amplifier

Many attempts have been made in the past to provide a low phase noise 10 MHz signal as a frequency reference, however recently I experienced some trouble because of ground loops. Normally no problem to decouple from DC voltages, but still the ground stays connected. The only way to avoid such ground loops is to use potential-free isolation, best using transformers. Capacitive coupling may be an option, but it is best avoided, at least it is though to get good isolation, say 2 kV or above, with capacitors that can transmit 10 MHz, at reasonable cost and size.

I am looking for about 1 V p-p, reasonably square shape output, into 50 Ohms, or TTL level (about 5 V) into high impedance. About 5-10 dBm at the 1st harmonic, 10 MHz. So we need to drive about 15 mA through a 50 Ohm load.

As amplifier elements, I am using 74HCU04 unbuffered inverters, these are balanced for propagation delay, and I have plenty of these in a box. The HCU04 is essentially a single stage inverter, a gate with a pretty good linear region – an amplifier. Propagation delay is about 5 ns at room temperature, so it is good solution to amplify clocks, and so on. We are using it to amplify a 10 MHz signal from an OCXO.

For isolation, looking for some small transformers (generally speaking ethernet transformers will work well), I found the PE-65612NL at low cost (list price is about 4 USD per piece, but some sellers have them at a small fraction of this cost, most likely, from surplus). These are 1:1, 2 kV min, signal transformers originally intended for digital audio signal separation. Good enough for our purposes.

A really affordable offer… sure you can substitute any other reasonable signal transformer that can cope with at least 20 mW, and is reasonably inexpensive.

The schematic – first, a single HCU04 is used to square up the OCXO output, and then distribute to 3 outputs, two are used to drive 2 isolated outputs each (4 outputs total), the other output is routed to a PLL circuit (because this isolation amp is part of a GPSDO). Any phase drift of the 1st stage HCU04 introduced by thermal and other slow effects will be canceled to some part by the GPS loop (because the sampling of the phase is very close to the isolated outputs, only followed by a set of paralleled-up gates) – although I don’t expect such drift.

The resistors were selected as 3×330 Ohm, giving about 100 Ohms source resistance and about 1.4 V pp when terminated in 50 Ohms.

Output power is fairly consistent, like, +-0.2 dBm when comparing 4 units. Fundamental output at 8 dBm is exactly the right range. Probably you can adjust it in the range of 5 to 10 nominal without changing much the other characteristics of the circuit, by changing the resistor values from the paralleled-up gates to the isolation transformer.

u-blox GPSDO: Joe’s and Gisela’s magic generator

In reply to an earlier post, GPSDO Update, I received the following great implementation of a GPSDO using u-blox receivers. The pictures are rather self explanatory.
>>>>
Hello again Simon!
I trust you are well and are enjoying the year end break.

We ( My good Wife and I..) have put your GPSDO software to good use. We used your message processing code almost as is, and added the various functions to drive my specific hardware and DAC, etc.

I have built up the complete GPSDO, with the 1-50MHz Analogue Devices AD9854 Quadrature DDS as a signal generator, provided with a 200MHz clock from the SI5351 PLL.
I also have a ‘signal generator’ output from another Si5351 channel, 1 to 200MHz, and a third channel output. square wave, from the GPS time pulse output, and can set outputs from 1Hz to 10MHz in decade steps.
I used a 7inch NEXTION graphics display for the display and control inputs ( touch screen) – that works very nicely!

I have run the unit for a few days now, and logged a 48hour period of data, every 10seconds, regarding the clock bias, drift, DAC output voltage etc, and the result looks very good indeed.
I am very pleased with the instrument and grateful for your assistance in providing your code. Thank You!

I have attached a few photos for interest.

Kind regards, and have a very good Christmas!

Joe and Gisela.
>>>>

TWS-N15 Noise Source 10 MHz-2 GHz: a few more sets

Coming back to an earlier post, Noise source design, I wanted to post the final results, and the looks of these noise sources.

The case is an aluminum extrusion design, and the lids are milled to accomodate the BNC and SMA connectors. The SMA is a really high quality connector. No point in using a noise source with a cheap connector – you are normally going to connect and disconnect this often.

The constant current supply is optimized for the maximum noise output, normally, about 8 mA. The design is a current mirror, with a TL431 precision reference.

The noise section is soldered with 0603 SMD mostly, on a FR4 board.

Foam and copper tape to avoid any foreign signals getting into it. Spurious signals can mean big trouble with noise measurements.

Return loss, I think it is pretty good.

SWR.

After some optimization of the circuit, the ENR output is now pretty flat, even with no specially expensive noise diode.

After all, pretty happy with the device, and others are happy two, as I give them away at low cost. If you need one, let me know.